Difference between revisions of "User:Shawndouglas/sandbox/sublevel10"

From LIMSWiki
Jump to navigationJump to search
(Replaced content with "<div class="nonumtoc">__TOC__</div> {{ombox | type = notice | style = width: 960px; | text = This is sublevel10 of my sandbox, where I play with features and...")
Tags: Reverted Replaced
(Undo revision 44006 by Shawndouglas (talk))
Tags: Undo Reverted
Line 3: Line 3:
| type      = notice
| type      = notice
| style    = width: 960px;
| style    = width: 960px;
| text      = This is sublevel10 of my sandbox, where I play with features and test MediaWiki code. If you wish to leave a comment for me, please see [[User_talk:Shawndouglas|my discussion page]] instead.<p></p>
| text      = This is sublevel2 of my sandbox, where I play with features and test MediaWiki code. If you wish to leave a comment for me, please see [[User_talk:Shawndouglas|my discussion page]] instead.<p></p>
}}
}}


==Sandbox begins below==
==Sandbox begins below==
==4. Results Review and Verification==
{|
| STYLE="vertical-align:top;"|
{| class="wikitable collapsible" border="1" cellpadding="10" cellspacing="0"
|-
  ! colspan="2" style="text-align:left; padding-left:20px; padding-top:10px; padding-bottom:10px;"|
|-
  ! style="color:brown; background-color:#ffffee; width:500px;"| Regulation, Specification, or Guidance
  ! style="color:brown; background-color:#ffffee; width:700px;"| Requirement
|-
  | style="padding:5px; width:500px;" |
[https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68 (b)]<br />
[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />
[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-4-1]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br />
[https://www.iso.org/standard/56115.html ISO 15189:2012 5.5.1.4]
  | style="background-color:white;" |'''4.1''' The system shall allow for the verification of specification limits for accuracy or tolerances while indicating to the user when values are exceeded.
|-
  | style="padding:5px; width:500px;" |
[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-4-2]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.2]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br />
[https://www.iso.org/standard/56115.html ISO 15189:2012 5.9.2]
  | style="background-color:white;" |'''4.2''' The system shall have the capability of automatically authorizing verification check tests and samples if they are within specification limits set by the instruments used to perform the tests.
|-
  | style="padding:5px; width:500px;" |
[https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68]<br />
[https://www.law.cornell.edu/cfr/text/21/211.100 21 CFR Part 211.100]<br />
[https://www.law.cornell.edu/cfr/text/21/211.160 21 CFR Part 211.160 (a)]<br />
[https://www.law.cornell.edu/cfr/text/21/211.188 21 CFR Part 211.188]<br />
[https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br />
[https://www.law.cornell.edu/cfr/text/45/170.315 45 CFR Part 170.315 (d)]<br />
[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-4-3]<br />
[https://www.fbi.gov/services/cjis/cjis-security-policy-resource-center CJIS Security Policy 5.4.1.1]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.1 and 4.9.7]<br />
[https://ec.europa.eu/health/sites/health/files/files/eudralex/vol-1/dir_2003_94/dir_2003_94_en.pdf E.U. Commission Directive 2003/94/EC Article 9.2]<br />
[https://extranet.who.int/prequal/content/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 15.7 and 15.9]
  | style="background-color:white;" |'''4.3''' The system shall accurately maintain a full audit trail for modified results.
|-
  | style="padding:5px; width:500px;" |
[https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68 (b)]<br />
[http://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. E-28–29]<br />
[[https://www.aihaaccreditedlabs.org/Policies/Pages/default.aspx AIHA-LAP Policies 2018 2A.7.7.2 and 7.7.3]<br />
[http://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 4.13.2.12]<br />
[http://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.9.4–5]<br />
[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-4-4]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.6]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.2–3 and 4.9.11]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.4]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br />
[https://ec.europa.eu/health/sites/health/files/files/eudralex/vol-4/annex11_01-2011_en.pdf E.U. Annex 11-6]<br />
[https://www.iso.org/standard/56115.html ISO 15189:2012 5.7.1]<br />
[https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.1.1]<br />
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Data and Instrumentation for PDP 10]
  | style="background-color:white;" |'''4.4''' The system shall provide one or more levels of review, as well as interpretation and documentation of results—whether entered manually or via an automated process—before release.
|-
  | style="padding:5px; width:500px;" |
[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-4-5]<br />
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br />
[https://ec.europa.eu/health/sites/health/files/files/eudralex/vol-4/annex11_01-2011_en.pdf E.U. Annex 11-6]<br />
[https://www.iso.org/standard/56115.html ISO 15189:2012 5.9.2]
  | style="background-color:white;" |'''4.5''' The system shall identify out-of-range results and clearly alert the appropriate individuals for further evaluation of the results.
|-
  | style="padding:5px; width:500px;" |
[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-4-6]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br />
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br />
[https://ec.europa.eu/health/sites/health/files/files/eudralex/vol-4/annex11_01-2011_en.pdf E.U. Annex 11-6]
  | style="background-color:white;" |'''4.6''' The system shall allow for the comparison of entered results with pre-defined specification limits from master data, clearly alerting the user when the results exceed those specification limits.
|-
|}
|}
==5. Sample, Experiment, and Study Approval and Verification==
{|
| STYLE="vertical-align:top;"|
{| class="wikitable collapsible" border="1" cellpadding="10" cellspacing="0"
|-
  ! colspan="2" style="text-align:left; padding-left:20px; padding-top:10px; padding-bottom:10px;"|
|-
  ! style="color:brown; background-color:#ffffee; width:500px;"| Regulation, Specification, or Guidance
  ! style="color:brown; background-color:#ffffee; width:700px;"| Requirement
|-
  | style="padding:5px; width:500px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-5-1]<br />[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sample Processing and Analysis Procedures for PDP 5.9–10]
  | style="background-color:white;" |'''5.1''' The system shall accurately record details of a sample's final disposition.
|-
  | style="padding:5px; width:500px;" |[https://www.aihaaccreditedlabs.org/Policies/Pages/default.aspx AIHA-LAP Policies 2018 2A.7.7.2]<br />[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-5-2]
  | style="background-color:white;" |'''5.2''' The system should allow authorized personnel to view relevant metadata for results during the review and approval process, including the instruments and reagents used, raw data from instruments, and associated reports.
|-
  | style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/21/211.160 21 CFR Part 211.160 (b)]<br />[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-5-3]<br />[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. IV, Sec. 6.1]
  | style="background-color:white;" |'''5.3''' The system should provide appropriate functions for specific sample statuses like "approved" and "rejected," including the ability to handle re-test, re-sampling, re-calculation, out-of-specification response and notification, and disposition functions.
|-
|}
|}

Revision as of 13:22, 6 September 2021

Sandbox begins below

4. Results Review and Verification

Regulation, Specification, or Guidance Requirement

21 CFR Part 211.68 (b)
42 CFR Part 493.1241
ASTM E1578-18 C-4-1
EPA ERLN Laboratory Requirements 4.9.11
EPA QA/G-5 2.4.1
ISO 15189:2012 5.5.1.4

4.1 The system shall allow for the verification of specification limits for accuracy or tolerances while indicating to the user when values are exceeded.

ASTM E1578-18 C-4-2
EPA ERLN Laboratory Requirements 4.9.2
EPA ERLN Laboratory Requirements 4.9.11
EPA QA/G-5 2.4.1
ISO 15189:2012 5.9.2

4.2 The system shall have the capability of automatically authorizing verification check tests and samples if they are within specification limits set by the instruments used to perform the tests.

21 CFR Part 211.68
21 CFR Part 211.100
21 CFR Part 211.160 (a)
21 CFR Part 211.188
21 CFR Part 211.194
45 CFR Part 170.315 (d)
ASTM E1578-18 C-4-3
CJIS Security Policy 5.4.1.1
EPA ERLN Laboratory Requirements 4.9.1 and 4.9.7
E.U. Commission Directive 2003/94/EC Article 9.2
WHO Technical Report Series, #986, Annex 2, 15.7 and 15.9

4.3 The system shall accurately maintain a full audit trail for modified results.

21 CFR Part 211.68 (b)
ABFT Accreditation Manual Sec. E-28–29
[AIHA-LAP Policies 2018 2A.7.7.2 and 7.7.3
ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 4.13.2.12
ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.9.4–5
ASTM E1578-18 C-4-4
EPA ERLN Laboratory Requirements 3.2.6
EPA ERLN Laboratory Requirements 4.9.2–3 and 4.9.11
EPA ERLN Laboratory Requirements 4.11.4
EPA QA/G-5 2.2.10
EPA QA/G-5 2.4.1
E.U. Annex 11-6
ISO 15189:2012 5.7.1
ISO/IEC 17025:2017 7.8.1.1
USDA Data and Instrumentation for PDP 10

4.4 The system shall provide one or more levels of review, as well as interpretation and documentation of results—whether entered manually or via an automated process—before release.

ASTM E1578-18 C-4-5
EPA ERLN Laboratory Requirements 4.9.11
EPA QA/G-5 2.2.10
EPA QA/G-5 2.4.1
E.U. Annex 11-6
ISO 15189:2012 5.9.2

4.5 The system shall identify out-of-range results and clearly alert the appropriate individuals for further evaluation of the results.

ASTM E1578-18 C-4-6
EPA QA/G-5 2.2.10
EPA QA/G-5 2.4.1
E.U. Annex 11-6

4.6 The system shall allow for the comparison of entered results with pre-defined specification limits from master data, clearly alerting the user when the results exceed those specification limits.

5. Sample, Experiment, and Study Approval and Verification

Regulation, Specification, or Guidance Requirement
ASTM E1578-18 C-5-1
USDA Sample Processing and Analysis Procedures for PDP 5.9–10
5.1 The system shall accurately record details of a sample's final disposition.
AIHA-LAP Policies 2018 2A.7.7.2
ASTM E1578-18 C-5-2
5.2 The system should allow authorized personnel to view relevant metadata for results during the review and approval process, including the instruments and reagents used, raw data from instruments, and associated reports.
21 CFR Part 211.160 (b)
ASTM E1578-18 C-5-3
EPA 815-R-05-004 Chap. IV, Sec. 6.1
5.3 The system should provide appropriate functions for specific sample statuses like "approved" and "rejected," including the ability to handle re-test, re-sampling, re-calculation, out-of-specification response and notification, and disposition functions.